Sterilization Process Indicator- Chemical Indicator
- Recalling firm
- KCI New Zealand Unlimited
- Product
- Sterilization process indicator- Chemical Indicator. [Device]. Affected: Catalogue Number: 1322-24MM, 1322-18MM - Lot number: 251106 (distributed between December 10, 2025 through June 4, 2026)
- Reason
- Through an investigation by the product support team, it was determined that this lot may contain rolls without indicator ink present due to a manufacturing printing nonconformance. If the indicator tape roll does not contain indicator ink, the product will not provide visual distinction between processed and unprocessed units. The volume of potentially impacted rolls within the identified lot is low.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-07-13
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory