Sterilization Process Indicator- Chemical Indicator

Recalling firm
KCI New Zealand Unlimited
Product
Sterilization process indicator- Chemical Indicator. [Device]. Affected: Catalogue Number: 1322-24MM, 1322-18MM - Lot number: 251106 (distributed between December 10, 2025 through June 4, 2026)
Reason
Through an investigation by the product support team, it was determined that this lot may contain rolls without indicator ink present due to a manufacturing printing nonconformance. If the indicator tape roll does not contain indicator ink, the product will not provide visual distinction between processed and unprocessed units. The volume of potentially impacted rolls within the identified lot is low.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-07-13
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/04e4564c-1415-44e3-96da-78a9661ac179