Boston Scientific Cork, Ltd.: Catheter, Continuous Flush
Recall Summary Boston Scientific is recalling Direxion Torqueable Microcatheters worldwide due to manufacturing defects in the coating and metal components that could affect device performance. The hydrophilic coating on the catheter tip may not have been properly cured, and the metal sleeves may have surface irregularities from laser cutting errors. The recall affects catheters distributed across the US, Puerto Rico, and numerous countries internationally.
- Recalling firm
- Boston Scientific Cork, Ltd.
- Product
- Direxion Torqueable Microcatheter, NGMC/SINGLE/027/STRAIGHT/1RO/105; Material Number: M001195400
- Reason
- The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-09-16
- Status
- Open, Classified
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide distribution US Nationwide and territory of Puerto Rico. The countries of Albania, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy,...