Chromphare Surgical Lights

Chromphare Surgical Lights Recall Chromphare surgical lights are being recalled and will be exchanged. This recall affects hospitals, pharmacies, and laboratories in New Zealand that received these devices. The recall is classified as Class I, which is the most serious classification level.

Recalling firm
Stryker New Zealand Ltd
Product
Chromphare surgical lights. [Device]. Model: Devices manufactured with Soffit Rings. Affected: Part number - CH00000001
Reason
The manufacturer has discovered that certain lots of the soffit ring may not adequately support the weight of the ceiling cover. As a result, the cover may shift or slide down the ceiling tube. Movement of the ceiling cover can expose the area above the ceiling, which may allow particles, such as dust, debris, residue, or other contaminants, to enter the sterile field.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-04-13
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory