Philips Medical Systems Nederland B.v. Veenpluis: Interventional Fluoroscopic X-ray System
- Recalling firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis
- Product
- Philips Azurion 7B20/15, Model Numbers: 722068, 722226 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
- Reason
- During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2023-12-07
- Status
- Open, Classified
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- US Nationwide distribution.