Isotis Orthobiologics, Inc.: Osteoballast, 45 X 17.5 Mm (seaspine Brand), Ref: 56800095. Ballast, 45 X 17.5 Mm (isotis Brand), Ref: 02-8000-095

Recall Summary IsoTis OrthoBiologics is recalling OsteoBallast and Ballast bone graft products (specific model sizes and reference numbers listed) used in spine and pelvis surgery because some units were filled with insufficient bone matrix material—about 4.5 cc instead of the required 7.5 cc. This underfill may result in delayed, incomplete, or failed bone fusion after surgery. The products were distributed nationwide across multiple states.

Recalling firm
IsoTis OrthoBiologics, Inc.
Product
OsteoBallast, 45 x 17.5 mm (SeaSpine Brand), REF: 56800095. Ballast, 45 x 17.5 mm (IsoTis Brand), REF: 02-8000-095
Reason
Demineralized bone matrix (DBM) in resorbable mesh indicated for use with autograft as a bone graft extender in the posterolateral spine and pelvis, were filled with approx. 4.5 cc DBM (fill for 45 ¿ 11 mm configuration) instead of approx. 7.5 cc DBM (fill for 45 ¿ 17.5 mm configuration), which may lead to delayed fusion, incomplete fusion, pseudarthrosis, or lack of bone ingrowth.
Classification
Class II
Category
medical device recall
Country
United States
Agency
U.S. Food and Drug Administration
Recall date
2026-07-02
Status
Ongoing
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the states of AK, NJ, TN, AR, CA, WA, OH.