Baxter Healthcare Corporation: Baxter Novum Iq Syringe Pump, Model Number 40800baxus

Recall Summary Baxter Healthcare is recalling the Novum IQ Syringe Pump (Model 40800BAXUS) due to a software problem that can occur when using multiple syringes in one infusion session. If a syringe empties and is replaced while medication remains to be delivered, the pump may not function correctly. The pump has been distributed nationwide across all U.S. states and territories.

Recalling firm
Baxter Healthcare Corporation
Product
Baxter Novum IQ Syringe Pump, Model Number 40800BAXUS
Reason
A software issue that may occur during a Volume/Time (VOT) multi-syringe infusion when a syringe reaches empty and is replaced while a Volume To Be Infused (VTBI) remains.
Classification
Class II
Category
medical device recall
Country
United States
Agency
U.S. Food and Drug Administration
Recall date
2026-08-11
Status
Ongoing
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.

Official recall notice from U.S. Food and Drug Administration

Recall details on Nibble Recalls: https://trynibble.app/recalls/001503ca-42d3-44ed-aa39-157c7f930e8b